What to check if there is no AC power supply to my POLLOGEN Medical Equipment unit?
M
mathew41Sep 10, 2025
If there is no AC power supply to your POLLOGEN Medical Equipment unit, one of the inlet fuses may be burned. Ensure that there is voltage in the power supply socket. If the ON/OFF switch is still not illuminated in the ON position, contact your local Lumenis representative for service.
L
Laura VaughanAug 23, 2025
What to do if the applicator of my POLLOGEN Medical Equipment does not emit an energy pulse?
B
Barry BanksAug 23, 2025
If the applicator of your POLLOGEN Medical Equipment does not emit an energy pulse to the treatment area, it might be due to the patient controlled manual switch being pressed. Release the patient controlled manual switch and the foot switch, then press the foot switch again. If the system still does not operate properly, contact your local Lumenis representative for service.
C
Chris JohnsonAug 25, 2025
What causes the applicator of my POLLOGEN Legend Pro Medical Equipment not to emit an energy pulse?
P
Patrick GarciaAug 25, 2025
If the applicator of your POLLOGEN Medical Equipment does not emit an energy pulse, the cause may be a foot switch failure. If the system still does not operate properly, contact your local Lumenis representative for service.
M
Meghan SmithAug 26, 2025
How to fix an applicator connection problem on my POLLOGEN Medical Equipment that prevents energy pulse emission?
H
harry93Aug 27, 2025
If the applicator of your POLLOGEN Medical Equipment is not emitting an energy pulse, check that the relevant applicator is properly connected with the locking mechanism. If it's still not working, replace the relevant applicator. If the applicator still does not operate properly, contact your local Lumenis representative for service.
W
William CarsonAug 28, 2025
What to do if the applicator on my POLLOGEN Medical Equipment does not heat the treatment area?
J
John BatesAug 28, 2025
If the applicator on your POLLOGEN Medical Equipment does not heat the treatment area, ensure that the applicator is connected to the correct connector. If the system still does not operate properly, contact your local Lumenis representative for service.
V
Valerie HillAug 29, 2025
Why is the applicator of my POLLOGEN Legend Pro not heating the treatment area?
J
Jeanne GibsonAug 29, 2025
If the applicator of your POLLOGEN Medical Equipment is not heating the treatment area, it may be because the patient controlled manual switch is not connected. Connect the patient controlled manual switch. If the system still does not operate properly, contact your local Lumenis representative for service.
D
Dale ThompsonSep 18, 2025
What could cause no energy output from my POLLOGEN Legend Pro?
T
thomasheatherSep 18, 2025
If there is no energy output from your POLLOGEN Medical Equipment, it may be due to a PPB failure. Contact your local Lumenis representative for service.
B
Brian HorneSep 12, 2025
What to do if my POLLOGEN Legend Pro system cannot be switched off?
M
Mary ChandlerSep 12, 2025
If your POLLOGEN Medical Equipment system cannot be switched off, the cause may be an ON/OFF switch failure. Unplug the device and contact your local Lumenis representative for service.
M
Melanie SanchezSep 8, 2025
What to do if the energy cannot be adjusted on my POLLOGEN Legend Pro Medical Equipment?
T
Tracy HardySep 8, 2025
If the energy cannot be adjusted on your POLLOGEN Medical Equipment, it may be due to a failure to enter treatment mode. Press shortly on FS to enable Mode calibration.
Y
yoderjessicaSep 5, 2025
What should I do if my POLLOGEN Legend Pro system doesn't stop treatment when the foot switch is released?
C
Collin NelsonSep 5, 2025
If your POLLOGEN Medical Equipment system does not stop treatment when the foot switch is released, it could be due to a foot switch failure. Switch the system OFF and then switch it ON again to check the foot switch. If the system still does not operate properly, contact your local Lumenis representative for service.