17. ELECTROMAGNETIC COMPATIBILITY TABLE SHOWN IN SECTION 5.2.2 IEC 60601-
1-2 Ed. 4.0 (2014)
Summary of Test Results
1.
Emission (IEC 60601-1-2 sec. 7, IEC 60601-2-2 sec. 202.6.1 & IEC 60601-2-10 sec. 202.6.1)
Conducted emission
Freq. range:150 kHz - 30 MHz
Group 1 Class B
230 &100 VAC mains
Radiated emission
Freq. range: 30 - 1000 MHz
Harmonic current emission
test
Voltage changes, Voltage
fluctuations and Flicker test
Immunity (IEC 60601-1-2 section 8, IEC 60601-2-2 sec. 202.6.2 & IEC 60601-2-10 sec. 202.6.2)
Immunity from Electrostatic
discharge (ESD)
8 kV contact discharges &
15 kV air discharges
Immunity from radiated
electromagnetic fields
3.0 V/m; 80 MHz 2.7 GHz,
80% AM, 1 kHz
Immunity from Proximity field
from wireless communications
equipment
List of frequencies,
from 9 V/m up to 28 V/m,
PM (18 Hz or 217 Hz), FM 1 kHz
Immunity from Electrical Fast
transient (EFT)
± 2.0 kV on AC mains;
Tr/Th – 5/50 ns, 100 kHz
±2.0 CM / ±1.0 kV DM on AC mains;
Tr/Th – 1.2/50 (8/20) s
Immunity from conducted
disturbances induced by
radio-frequency fields
3.0; 6.0 VRMS on AC mains,
& Applicators cables;
0.15÷ 80 MHz, 80% AM 1 kHz
Immunity from power
frequency magnetic field
Immunity from voltage dips,
short interruptions and
voltage variations
On AC mains:
0 % - 0.5 cycle & 1 cycle;
70% - 25 cycles; 0% - 250 cycles