Legend Pro System User Manual Version 7
Proprietary Information of Pollogen a company of Lumenis 9
Do not treat the area near the thorax which may increase the risk of cardiac fibrillation.
Never treat over the eye socket, the chest, breasts or underarms.
Never treat over the thyroid area.
Patient should not come into contact with earthed metal parts or parts with appreciable capacitance to
earth.
Regularly inspect the accessories for possible damage; in particular, electrode cables and HF energized
devices.
2.4 VO and disposable related warnings :
Never direct the Applicator’s tip at anything other than the treatment site.
Reuse of single-use disposable tips creates a potential risk of self or cross-contamination and infection
which may each lead to significant medical complications. Do not treat another patient with the same
tip.
Do not allow the Applicator to come in contact with hard materials that could damage the tip.
Place the tip perpendicular to the skin with full contact, applying slight pressure.
Ensure that the Applicator’s tip is free of debris and thoroughly dry.
No lotion, gel, oil or glycerin is used when treating with the VO Applicator. Verify clean & dry skin.
Always discard the tip at the end of patient’s treatment as biological hazardous waste.
Do not use a tip if its expiry date is overdue.
Do not use a tip if its package is compromised. Verify package integrity before use.
Always check the integrity of the tip. Verify that all the electrodes are present. If any pins are missing, do
not use the tip.
Ensure that the tip is clean during treatment. Accumulated debris or non-dry skin may cause pain and
skin damage.
During treatment inspect the tip’s integrity and cleanliness once every 10 to 15 pulses. This is especially
important when using high parameters, which may cause increased buildup of debris. If the tip is not
clean it must be replaced.
After VO treatment, no other treatment should be performed until the treated area has completely
healed.
To protect from unwanted contact of the electrode pins, make sure to secure the tip using its protective
cover when not in use.
2.5 Electrical and Mechanical Safety warnings :
High voltage is present inside the Legend Pro™ System. Always take proper precautions as described in
this manual, while making sure to constantly be aware of the possible dangers.
Only qualified personnel that have been trained by Lumenis® are allowed to supply service to this
equipment.
Do not touch the inner parts of the system; service is supplied only by company’s authorized personnel.
To avoid risk of electric shock, this equipment must only be connected to a supply main with protective
earth and must be used with the cable supplied. (cable10A, 250V, 1.8m length). The system is grounded
through the grounding conductor in the power cable. This protective grounding is essential for safe
operation.
Keep all covers and panels of the System securely closed at all times when in use. Removal of the covers
or panels can create a safety hazard.
Perform maintenance procedures when the System is shut down and disconnected from power outlets.
Always keep the System’s wheel breaks locked to avoid uncontrolled movement. When movement is
necessary, unlock the breaks before doing so. Always move the System slowly and carefully. The System
may cause injury if proper care is not taken when moving it.

Table of Contents