Legend Pro™ System User Manual Version 7
Proprietary Information of Pollogen a company of Lumenis 57
No text on the
TOUCH SCREEN
ON/OFF power switch is lit.
Switch OFF the system and then
switch ON again.
If there is still no display on
TOUCH SCREEN, stop system
operation and contact local
Lumenis representative for
service.
No backlight on
the TOUCH
SCREEN
AC/DC Power
supply failure
No backlight on the touch
screen.
ON/OFF power switch is lit.
No indication on either
Applicator.
Switch OFF the system and then
switch ON again.
If there is still no display on
TOUCH SCREEN, stop system
operation and contact local
Lumenis representative for
service.
Partial data displayed on the
touch screen.
Switch OFF the system and then
switch ON again.
If there is still no display on
TOUCH SCREEN, stop system
operation and contact local
Lumenis representative for
service.
System does not
stop treatment
when Foot
Switch is
released
Screen display still operating
when Foot Switch is released.
Switch OFF the system and then
switch ON again to check Foot
Switch.
If System still does not operate
properly, contact local Lumenis
representative for service.
No AC power
supply to the unit
One of the
inlet fuses is
burned
No display on the touch
screen.
ON/OFF switch is not
illuminated in ON position.
Ensure that there is voltage in
power supply socket.
If ON/OFF switch is still not
illuminated in ON position –
contact local Lumenis
representative for service.
System cannot
be switched off
System cannot be switched
off.
Unplug the device.
Contact local Lumenis
representative for service.
Unit Fan is not
operating
Fan or fan
connection
failure.
Contact local Lumenis
representative for service.
Touch screen
does not operate
as a control
panel
Touch screen not operating.
Contact local Lumenis
representative for service.