6 Working with the Application Menu
54
6.12 Setup for the statistics-program
See Chapter 4.4 “Activating the two main menus” and Chapter 4.5
“How the menu control operates” to activate the Application menu.
Functions of the statistics program and storage functions:
•MODE
In this function-field, you define, whether only the statistics-pro-
gram, only the storage-program or both programs simultaneously
should be used.
•RECORDING
With “MANUAL”, the «STO» function key must be pressed for
each value to be stored.
With “LOADCHANGE” the balance stores the measured value au-
tomatically after a load change.
With “TIMEBASE” the balance stores every value measured after
a defined period (works setting: 2.0 seconds).
TIMEBASE
Definition of the time span for the recording of data in accordance
with “RECORDING TIMEBASE” (for example, every 2 seconds).
•SET
STATISTIC
STATISTIC MODE OFF
MODE STATISTIC
MODE RECORDER
MODE STAT./RECORDER
Statistics program off
Statistics only
Data storage only
Stat. and storage
COUNT 100 Number of values to be
stored automatically
(1...999)
RECORDING MANUAL With «STO» function
key
RECORDING TIMEBASE On a time basis
RECORDING LOADCHANGE After every weight
change
TIMEBASE 2.0 Time base for “storage”
in seconds