■ 6 Working with the Application Menu
56
• Display-sequence, if changed with «
D»:
6.13 Setup for Check-Weighing
See Chapter 4.4 “Activating the two main menus” and Chapter 4.5
“How the menu control operates” for the activation of the Application
menu.
With the aid of the «CHECK +/-» program you can check each mea-
surement for its agreement with a defined reference-value plus/minus
allowable deviations.
In the «CHECK +/-» application the four function keys are not active.
In the display “+”, “–” and «
ÆIIÅ» are active.
If «
ÆIIÅ» lights up, the measured value lies within the specified
tolerances.
+ 123.456 9
Display sequence:
MEAN: 123.456 g
STDE.: 0.001 g
STDE.-%: 0.01 %
MAX: 123.457 g
MIN: 123.456 g
„individual values“
Exit with «esc»
MEAN 123.456 9
GGGG
«A»«B»«
C»«D»
then:
+ 123.456 9
STDE 0.001 9
GGGG
«A»«B»«
C»«D»
• SET CHECK +/-
CHECK +/- MODE ON/OFF
NOM. 100.000 g
TO 120.000 g
TU 80.000 g
Switch application on/off
Enter nominal weight
Define over load
Define under load
NOTE
As an accessory, a signal light is available for this type of display.
See Chapter 13 “Accessories”.