14 Further information
82
Then proceed as follows:
14.2.3 Internal calibration
For an internal calibration with the built-in calibration weight “SET CA-
LIBRATION MODE INTERNAL” must be selected in the Configuration
menu (see Chapter 5.6 “Calibration functions”).
Then proceed as follows:
Switch to “BALANCING” with the Change key
•Press «T» until “CALIBRATION” is shown.
The calibration is finished after a certain period of time.
Switch to “BALANCING” with the
Change key
cALIBRation
•Press «T» until
“CALIBRATION” appears.
- - 0000 9
The balance carries out a Zero
measurement (0000 g is shown
flashing)
- - 372 9
After the zero measurement the
display flashes with the previous-
ly entered calibration weight
Place the calibration weight on
the pan
- - 372 9
The display flashes rapidly
Calibration is complete when the
display stops flashing (the exact
weight is shown)
+ 372.15 9