14 Further information ■
79
14 Further information
14.1 Notes on the Weighing mode
14.1.1 Set Weighing mode: FLOATINGDISPLAY
The value set for FLOATINGDISPLAY defines the period, after which
each new measurement is displayed.
For the definition of this period, the quality of the balance-location is
crucial. The stability control must also be suitably chosen.
Recommended values:
14.1.2 Set Weighing mode: Stability control
The value set for the stability control depends on the quality of the ba-
lance location and must be correctly chosen in order to obtain optimal,
reproducible results. Choose:
• “HIGH STABILITY” at an optimal balance location,
• “MEDIUM STABILITY” at a good balance location or
• “LOW STABILITY” at a critical balance location
• Optimal balance location: FLOATINGDISPLAY 0.04 or
FLOATINGDISPLAY 0.08
• Good balance location: FLOATINGDISPLAY 0.16
• Critical balance location: FLOATINGDISPLAY 0.32
NOTE
The value of the Floating Display is a function of the stability con-
trol and the balance location. For balance location, see Chapter 3.4
“Choice of a suitable location” and Chapter 5.7 “Weighing mode”.