8 Data transfer to peripheral devices
63
Tab. 8.4 Remote control-commands
8.4 Examples for the remote control of the
balance
Tab. 8.5 Examples for remote control
Sn Set stability n n=0 low
n=1 medium
n=2 high
SDTttmmjj
hhmmss
Set date and time (German) (Tag, Monat, Jahr,
Stunde, Minute, Sekunde)
SDTmmdd
yyhhmmss
Set Date and Time (English) (Month, Day, Year,
Hour, Minutes, Seconds)
T (ttt) Tare or set tare to a specific value
Uxnn
Set unit x (1...4) of the balance with nn (0=g,
1=mg, 2=kg, ...)
UxS Switch balance to unit x (1...4)
ZERO Zero balance (provided weight is stable and within
the zero position range)
NOTE
Each remote control-command must terminate with «CR» «LF».
The commands are acknowledged if required.
Input Description of the function executed
D – – – – – Five dashes will be shown
DTEST123 Will show: tESt123
D The display will be dark
T100 -100.000g (Tare set to 100 g)
T1 -1.00000 g (Tare set to 1 g)
T Balance will be tared
Command Function

Table of Contents