EasyManuals Logo

QuadTech Guardian 2500 Series User Manual

Default Icon
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
Page 26 of 26
Condensed Operating Instructions (continued)
PASS FAIL
3. Test Results Display with Single Test Selected (1 – 25)
If limits are not exceeded during the test the instrument will cycle through the complete test
and the Pass indicator (green) will light. The unit is ready for another test.
If limits are exceeded during the test the instrument’s high voltage will shut down, the
display will indicate the failed limit and an audible alarm will sound (only if buzzer is
enabled, see Utility function, paragraph 3.4). Press the STOP button two times, one to stop
the buzzer (if enabled) and again to reset before another test can be initiated.
One of the following “Display messages” will be shown (upper right of the display)
Display Test Condition
Pass Measurement falls within set limits
Lo Ramp Measurement was below low limit during ramp time
Hi Ramp Measurement exceeded high limit during ramp time
Lo Fail Measurement was below low limit during test time
Hi Fail Measurement exceeded high limit during test time
Arc Fail Measurement exceeded arc limit
GC Fail Measurement exceeded ground continuity high limit
NOTE
An asterisk following the parameter as illustrated in the measurement display above
(AC Tot *) indicates that the Measure Offset is selected for ON. No asterisk will be
shown in the measurement result with Measure Offset to OFF. Refer to Measure
Offset, paragraph 2.4.2
An asterisk following a Pass message as illustrated in the measurement display above
(Pass*) indicates a conditional pass. A conditional pass can occur when the enhanced
feature “Continue Voltage On Fail” is enabled. The asterisk indicates that the device
failed at some point during the test but was passing at the end of the test. This would
only be seen with “Continue Voltage On Fail” enabled. Refer to Continue Voltage On
Fail, paragraph 2.4.9, step 3.
1 0.325KV Pass*
AC Tot * 0.005mA 0.0s
1 0.325KV Hi Fail
AC Tot * 0.005mA 3.5s

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the QuadTech Guardian 2500 Series and is the answer not in the manual?

QuadTech Guardian 2500 Series Specifications

General IconGeneral
BrandQuadTech
ModelGuardian 2500 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals