Page 62 of 62 Operation/Programming
8. Tests To View
The following selection determines which tests in memory, when selected by test number, are
displayed in View mode. Single step are 1 through 25 and sequence are S1 through S9.
Press the UTILITY key and right arrow key to select Enhanced Features.
If Enhanced Features is set for On, press right arrow key to select Tests to View.
If Enhanced Features is set for Off, press up/down arrow key to select On and press
right arrow key to select Tests to View
Press up/down arrow keys to select Tests To View, All Tests (1-25, and S1 - S9),
Only Single Step (1 - 25), Only Sequence (S1 - S9).
Press the UTILITY key to exit or move to other Enhanced Features with the right
arrow key
9. Perform Calibration
This selection indicates the instruments expected calibration date and can be used to enter a
calibration routine. This procedure should only be used by qualified calibration personnel and
does require proper calibration equipment and password entry. Refer to the Service and
Calibration section of this manual for addition information.
Tests To View Util
All Tests
Tests To View Util
Only Sequence
Tests To View Util
Only Single Step