2.4 Utility Functions
2.4.1 General
The Utility functions allow the user to set the Measure Offset function for a currently selected
test and set other conditions that affect all tests. When the UTILITY key is pressed the first
function, Meas Offset, is displayed and the test number to which this applies is shown in the
upper left display. Other functions are accessed by pressing the right arrow key.
Table of Programmable Parameters For Utility Functions
Programmable
Functions
Settings
Measure Offset
On, Off,
Press Start-Get Offset
This function applies to a single test,
other functions apply to all tests
Keypad Lockout
Lock Password, Off
Lock Cycle Tests
Test Locations
Hide Disabled Tests
Show All Tests
Note: Bold denotes normal default
conditions when shipped from
Soft, Off, Loud
factory.
Beep on Key Hit
On, Off
RS232 Interface
9600, 19200, Disable
IEEE-488 Interface
(optional)
1 to 16
Enhanced Features
On, Off
Programmable
Functions With (On)
Settings
Test Low Limit
Continuous, End of
Test
Delay Time
Off, 0.1 to 999.9sec
Continue Voltage On
Fail
On, Off
applies only to a sequence test (S1-S9) -
Continue Steps On Fail
On, Off
applies only to an IR test -
Stop Test On Pass
On, Off
applies only to a sequence test (S1-S9) -
Step And Increment
Off, On
Clear All Tests
No, Yes
Tests To View
All Tests
Only Single Step
Only Sequence
Perform Calibration
(Follow procedure as
indicated on display)
Ground Fault Stop
Off, On
Off-DUT Grounded
Arc Pulse Width
1 to 50µsec
does not apply to the 2510 -
Adaptive DC Ramp
Off, On
Autoranging
Full, None, Limited