Opera
tion/
Prog
ramm
ing
Test Par
ameter Se
tup
NOTE
Some of the setups (21 -25, depending on model type) contain factory set test conditions. These
can be modified like all other setups (1-20) but can be returned to the original factory set
conditions with
Clear All Tests
, refer to paragraph 2.4.9, step 7.
Step
Test
Parameter
AC Hipot
(2510,2520,2530)
DC Hipot
(2520,2530)
IR
(2530)
GC
(2510,2520,2530)
1
PROGRAM
PROGRAM
PROGRAM
PROGRAM
2
Mode
AC Total Current
AC Real Current
AC Imag Current
DC Current
IR
GC Only
3
Test
Voltage
0.100KV to
5.000KV
0.100KV to
6.000KV
0.050KV to
1.000KV
4
Hi Limit
0.001mA
to 15mA
0.0001mA
to 7.5mA
0.01M
Ω
to
2.000T
Ω
, Off
0.01 to 10.00
Ω
5
Lo Limit
0.001mA to
14.999mA, Off
0.0001mA to
7.4999mA, Off
0.01M
Ω
to
1.999T
Ω
0.01 to 9.99
Ω
Off
6
Arc Limit
0.5mA to 15mA
Off
0.5mA to 15mA
Off
7
Ramp Time
0.1 to 999.9sec
Off
0.1 to 999.9sec
Off
0.1 to 999.9sec
Off
8
Hi Limit
Ramp
0.001mA to
15.000mA, Off
0.0001 to
7.500mA, Off
0.01M
Ω
to
2.000T
Ω
, Off
9
Lo Limit
Ramp
0.001mA to
14.999mA, Off
0.0001 to
7.499mA, Off
0.01M
Ω
to
1.999T
Ω
, Off
10
Dwell Time
0.1 to 999.9sec
Off
0.1 to 999.9sec
Off
11
Test Time
0.1 to 999.9sec
Continuous
0.1 to 999.9sec
Continuous
0.1 to 999.9sec
Continuous
0.1 to 999.9sec
Continuous
12
Fall Time
0.1 to 999.9sec
Off
0.1 to 999.9sec
Off
0.1 to 999.9sec
Off
13
Frequency
50 or 60Hz
14
Gnd
Continuity
0.01
to
10.00
Ω
Off
0.01
to
10.00
Ω
Off
0.01
to
10.00
Ω
Off
39
41
Table of Contents
Introduction - Section
1
Table of Contents
3
Warranty
7
Specifications
9
Accessories
13
Safety Precautions
17
Condensed Operating
19
Recalling Setups/Preprogrammed Test Conditions
20
Modifying/Programming Test Conditions
21
Implementing Measure Offset
23
Connection Using Test Leads
24
Connection Using Corded Product Adapter
24
Connection to S50 Using Corded Product Adapter
25
Initiating Tests
25
Unpacking/Inspection
29
Product Overview
29
Front Panel Description
30
Rear Panel Description
34
Installation
35
Instrument Positioning
35
Cleaning
36
Power Requirements
36
Safety Inspection
36
Fuse Replacement
37
Operation/Programming - Section 2 2.1 Power up
39
Program/Setup Procedure
39
Hi Limit
40
Lo Limit
40
Mode
41
Voltage
41
Arc Limit
43
Ramp Time
43
Hi Limit Ramp
43
Lo Limit Ramp
44
Dwell Time
44
Test Time
44
Fall Time
45
Frequency (AC Hipot Only, Model 2510/2530)
45
Ground Continuity
45
Programming Sequence Tests
46
Sequence Test #1
46
Sequence Test #2
47
Sequence Test #3
48
Calculate Measurement Function
49
Utility Functions
50
General
50
Measure Offset
51
Keypad Lockout
51
Test Location
53
Buzzer on Fail
54
Beep on Key Hit
54
RS-232 Interface (Optional)
55
IEEE-488 Interface (Optional)
55
Enhanced Features
56
Test Low Limit
56
Delay Time
57
Continue Voltage on Fail
58
Continue Steps on Fail
58
Stop Test on Pass
59
Step and Increment
60
Clear All Tests
60
Test to View
62
Perform Calibration
62
Ground Fault Stop
63
Arc Pulse Width
64
Adaptive DC Ramp
64
Autoranging
65
Default Test Conditions
67
General
67
Setups
67
Error Messages
68
Interface - Section 3 3.1 General
73
Remote I/O
73
Enhanced Remote I/O Interface
76
IEEE-488.2 Interface
78
General
78
IEEE-488 & Rs-232 Commands
79
Remote Calibration
87
Formats
88
IEEE-488 Sample Program (Quick Basic)
89
General
91
Instrument Return
91
Calibration
91
5
Based on 1 rating
Ask a question
Give review
Questions and Answers:
Need help?
Do you have a question about the QuadTech Guardian 2500 Series and is the answer not in the manual?
Ask a question
QuadTech Guardian 2500 Series Specifications
General
Brand
QuadTech
Model
Guardian 2500 Series
Category
Test Equipment
Language
English
Related product manuals
QuadTech Sentry 20 Plus
85 pages
QuadTech Sentry 30 Plus
85 pages