Operation/Programming Page 67 of 67
2.5 Default Test Conditions
2.5.1 General
2.5.2 Setups (21 thru 25)
S25 - UL, Appliances
S24 - EN, Appliances
(with continuity test)
S23 - EN, Continuity Test
(low current)
S22 - EN & UL, Information
Technology Equipment
S21 - Insulation Resistance (IR)
Demo
Setup (25)
Mode: AC Total Current
Volt: 1.250KV
Hi Limit: 5.000mA
Low Limit: Off
Arc Limit: Off
Ramp Time: 0sec
Hi Limit Ramp: Off
Low Limit Ramp: Off
Test Time: 1.0sec
Fall Time: Off
Frequency: 60Hz
Gnd Continuity: Off
Setup (24)
Mode: AC Total Current
Volt: 1.500KV
Hi Limit: 5.000mA
Low Limit: Off
Arc Limit: Off
Ramp Time: 0sec
Hi Limit Ramp: Off
Low Limit Ramp: Off
Test Time: 1.0sec
Fall Time: Off
Frequency: 50Hz
Gnd Continuity 1.00Ω
Setup (23)
Mode: GC Only
Hi Limit: 0.1Ω
Low Limit: Off
Test Time: 1.0sec
Setup (22), 2520 & 2530 only
Mode: DC Current
Volt: 2.150KV
Hi Limit: 0.5mA
Low Limit: Off
Arc Limit: Off
Ramp Time: 1.0sec
Hi Limit Ramp: Off
Low Limit Ramp: Off
Dwell Time: Off
Test Time: 1.0sec
Fall Time: 1.0sec
Gnd Continuity Off
Setup (21), 2530 only
Mode: IR
Volt: 0.500KV
Hi Limit: Off
Low Limit: 0.10MΩ
Ramp Time: 5.0sec
Hi Limit Ramp: Off
Low Limit Ramp: 0.01MΩ
Dwell Time: 2.0
Test Time: 5.0sec
Fall Time: Off
Gnd Continuity Off