QuadTech Guardian 2500 Series - Default Test Conditions; General; Setups

Default Icon
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation/Programming Page 67 of 67
2.5 Default Test Conditions
2.5.1 General
2.5.2 Setups (21 thru 25)
S25 - UL, Appliances
S24 - EN, Appliances
(with continuity test)
S23 - EN, Continuity Test
(low current)
S22 - EN & UL, Information
Technology Equipment
S21 - Insulation Resistance (IR)
Demo
Setup (25)
Mode: AC Total Current
Volt: 1.250KV
Hi Limit: 5.000mA
Low Limit: Off
Arc Limit: Off
Ramp Time: 0sec
Hi Limit Ramp: Off
Low Limit Ramp: Off
Test Time: 1.0sec
Fall Time: Off
Frequency: 60Hz
Gnd Continuity: Off
Setup (24)
Mode: AC Total Current
Volt: 1.500KV
Hi Limit: 5.000mA
Low Limit: Off
Arc Limit: Off
Ramp Time: 0sec
Hi Limit Ramp: Off
Low Limit Ramp: Off
Test Time: 1.0sec
Fall Time: Off
Frequency: 50Hz
Gnd Continuity 1.00Ω
Setup (23)
Mode: GC Only
Hi Limit: 0.1Ω
Low Limit: Off
Test Time: 1.0sec
Setup (22), 2520 & 2530 only
Mode: DC Current
Volt: 2.150KV
Hi Limit: 0.5mA
Low Limit: Off
Arc Limit: Off
Ramp Time: 1.0sec
Hi Limit Ramp: Off
Low Limit Ramp: Off
Dwell Time: Off
Test Time: 1.0sec
Fall Time: 1.0sec
Gnd Continuity Off
Setup (21), 2530 only
Mode: IR
Volt: 0.500KV
Hi Limit: Off
Low Limit: 0.10MΩ
Ramp Time: 5.0sec
Hi Limit Ramp: Off
Low Limit Ramp: 0.01MΩ
Dwell Time: 2.0
Test Time: 5.0sec
Fall Time: Off
Gnd Continuity Off