Operation/Programming Page 45 of 45
11. To set Fall Time:
Press up/down arrow keys to increase or decrease the fall time between Off and
0.1 to 999.9sec. During fall time the programmed voltage decreases linearly to zero. No
Hi/Lo limits are checked during this time.
Press right arrow key to move to the next parameter ( left arrow for previous
parameter) or PROGRAM key to exit program mode.
Note: Under failed conditions for high limit, low limit or arc limit the 2500 will still
execute the fall time (if programmed) during the shutdown process.
12. To set Frequency:
Press up/down arrow keys to toggle the frequency between 50 or 60Hz
Press right arrow key to move to the next parameter ( left arrow for previous
parameter) or PROGRAM key to exit program mode.
13. To set Gnd Continuity:
Press up/down arrow keys to increase or decrease the resistance limit between Off and
0.01 to 10.00Ω. The ground continuity measurement is made between the G.C. and Return
connections. A ground continuity test is performed prior to a Hipot of IR test when the limit
is set to a value other than Off. Any measured value above this limit is considered a
FAIL, no low limit check is performed.
Press right arrow key to move to the next parameter ( left arrow for previous
parameter) or PROGRAM key to exit program mode.