Page 60 of 60 Operation/Programming
6. Step And Increment
This selection determines how the unit will step through a sequence test (S1 - S9). When On is
selected, and a test initiated with the START button, the first step performed is the one currently
viewed within a sequence. The next step in a sequence is initiated only when the START button
is pressed again, and so on through a maximum of 3 steps. The sequence then starts again with
step 1. This feature only applies to sequence test setups (S1 through S9).
Press the UTILITY key and right arrow key to select Enhanced Features.
If Enhanced Features is set for On, press right arrow key to select Step And Increment.
If Enhanced Features is set for Off, press up/down arrow key to select On and press
right arrow key to select Step And Increment.
Press up/down arrow keys to select Stop Test On Pass, On or Off
Press the UTILITY key to exit or move to other Enhanced Features with the right
arrow key.
7. Clear All Tests
This selection clears all stored test setups. Viewing of test setups after clearing would indicate
user defined ones (1 - 20) as No Test Programmed and sequence tests (S1 ā S9 as None. User
setups 21 thru 25 are returned to factory set conditions, refer to paragraph 2.5.
Press the UTILITY key and right arrow key to select Enhanced Features.