Page 44 of 44 Operation/Programming
8. To set Lo Lim Ramp (AC or DC Current only, not available with Ramp Time set
Off):
Press up/down arrow keys to increase or decrease the limit setting between Off and
0.001mA and 14.999mA (for AC) or 0.0001mA and 7.499mA (for DC) or 0.01MΩ and
1999000MΩ (for Insulation Resistance).. This Lo Lim Ramp limit is checked only
during the programmed ramp time.
Press right arrow key to move to the next parameter ( left arrow for previous
parameter) or PROGRAM key to exit program mode.
9. To set Dwell Time (DC Current or Insulation Resistance only)
Press up/down arrow keys to increase or decrease the dwell time between Off and
0.1 to 999.9sec (for DC Current and Insulation Resistance). No Hi/Lo Limits are checked
during the programmed dwell time.
Press right arrow key to move to the next parameter ( left arrow for previous
parameter) or PROGRAM key to exit program mode.
10. To set Test Time:
Press up/down arrow keys to increase or decrease the test time between Continuous and
0.1 to 999.9sec. Both Hi and Lo current limits (if programmed) are checked during the
test time. When the test time is programmed for Continuous the high voltage remains
On until a failure occurs or the STOP button is pressed.
Press right arrow key to move to the next parameter ( left arrow for previous
parameter) or PROGRAM key to exit program mode.