COMPUTER INTERFACE
5-13
Operation Mode:
There are two operation modes the user can set: Hygiene
and Survey mode. Hygiene mode is a continuous
measurement mode. Survey mode is a manual start and
stop measure mode. See chapter 2 for more details on the
Hygiene and Survey mode.
Warm-up Time:
This option allows the user to choose the warm-up time of
the instrument, in seconds.
Comfort Beep:
This option allows the user to specify a time interval that
the buzzer will beep once to remind the user that the
monitor is on. If zero is entered for the time interval, then
the comfort beep feature is disabled.
Back Light:
This option allows the user to enter a time out period so
that the back light will be turned off automatically after
the time out period.
Clock from PC:
This option allows the PC clock to be downloaded to the
MiniRAE 2000 monitor so that the user does not need to
manually set the clock in the MiniRAE 2000 monitor.
Alarm Mode:
This option allows the alarm to be either latched or auto-
reset. Selecting the auto-reset option allows the monitor
to turn off alarms when the alarm situation is no longer
exist. The latched option allows the alarm to stay on as
long as the alarm situation is no longer exist and a key is
pressed to acknowledge the alarm situation.