EB200 Manual MEMory subsystem
4052.2000.02 4.44 E-7
. :CONTents? <name>|RX
Query of contents of memory location
Parameters:
<name> MEM0 to MEM999 | RX | CURRENT
Result:
Depending on the setting by the command FORMat:MEMory either an ASCII data set or a binary data
set is output:
The ASCII data set has the following structure:
<F>,<THR>,<DEM>,<BW>,<ANT>,<ATT>,<ATTA>,<SQUC>,<AFC>,<ACT>
<F> frequency (see SENS:FREQ:CW?)
<THR> squelch threshold (see OUTP:SQU:THR?)
<DEM> demodulation type (see SENS:DEM?)
<BW> bandwidth (see SENS:BWID?)
<ANT> antenna number (see ROUT:CLOS:STAT?)
<ATT> attenuator (see INP:ATT:STAT?)
<ATTA> attenuator auto (see INP:ATT:AUTO?)
<SQUC> squelch function (see OUTP:SQU:STAT?)
<AFC> AFC function (see (SENS:FREQ:CW:AFC?)
<ACT> set/reset for scan (1/0)
The binary data set is transferred as a <Definite Length Block> and has to be interpreted
according to the above-mentioned format.
Notes:
• During query of the receiver data set (RX) the parameter <ACT> is not defined and has to be
ignored.
• When trying to read out an empty memory location the error message "MEMORY EMPTY" is
generated.
Example:
MEMory:CONTents? MEM1 -> 98500000,34,FM,100000,#14(@1),0,1,1,0,1
. . :MPAR <name>,<ACT>
Setting the memory location parameter (MPAR = MemoryPARameter) <ACT>
Parameters:
<name> MEM0 to MEM999 | CURRENT
<ACT> setting/resetting the memory to the scan (ON/OFF or 1/0)
Example:
MEMory:CONTents:MPAR MEM1, OFF