EB200 Manual SENSe subsystem FREQuency
4052.2000.02 4.72 E-7
. . :SPAN<numeric_value>|UP|DOWN |MINimum|MAXimum
Selection of frequency range with option IF panorama.
The following ranges are available in addition to the IF bandwidths: 25, 50, 100, 200, 500, 1000 kHz. If
a frequency range is entered it will be brought up to the next higher range.
Parameters:
<numeric_value> frequency range
UP|DOWN taking the range after|before the current bandwidth
MINimum|MAXimum setting the minimum|maximum frequency range
*RST state:
1000 kHz
Example:
FREQuency:SPAN 25 kHz
. . :SPAN? [MINimum|MAXimum]
Query of frequency range with option IF panorama.
Parameters:
none query of current frequency range
MINimum|MAXimum query of minimum|maximum frequency range
Result:
Frequency value Hz
Example:
FREQuency:SPAN?
25000
. . :STARt <numeric_value>|MINimum|MAXimum
Setting of start frequency of a frequency scan
Parameters:
<numeric_value> frequency
MINimum|MAXimum setting the lowest|highest start frequency
*RST state:
20 MHz
Example:
FREQuency:STARt 118 MHz