EB200 Manual SENSe subsystem FREQuency
4052.2000.02 4.73 E-7
. . :STARt? [MINimum|MAXimum]
Query of start frequency of a frequency scan
Parameters:
none query of current start frequency
MINimum|MAXimum query of lowest|highest start frequency
Result:
Frequency in Hz
Example:
FREQuency:STARt? -> 118000000
. . :STEP
. . . [:INCRement] <numeric_value>|MINimum|MAXimum
Setting the stepwidth for the command SENSe:FREQuency[:CW|FIXed] UP|DOWN
Note:
For SSB demodulation (CW, LSB and USB), the frequency stepwidth will be set to 1 Hz.
Parameters:
<numeric_value> frequency stepwidth
MINimum|MAXimum setting the smallest|largest stepwidth
*RST state:
1 kHz
Example:
FREQuency:STEP 25 kHz
. . . [:INCRement]? [MINimum|MAXimum]
Query of stepwidth
Parameters:
none query of currently set stepwidth
MINimum|MAXimum query of smallest|largest stepwidth
Result:
Frequency stepwidth in Hz
Example:
FREQuency:STEP? -> 25000