EB200 Manual OUTPut subsystem
4052.2000.02 4.52 E-7
. . :CONTrol ONLY|WITHaf
It can be selected whether, in the TONE mode, only the level tone or also the AF is output via the audio
channel.
Parameters:
WITHaf level tone and AF is output.
ONLY only level tone is output.
*RST state:
ONLY
Example:
OUTPut:TONE:CONTrol ONLY
. . :CONTrol?
Query of whether, in the TONE mode, only the level tone or also the AF is output via the audio channel.
Parameters:
none
Result:
ONLY, WITH
Example:
OUTPut:TONE:CONTrol? -> ONLY
. . :THReshold <numeric_value>|UP|DOWN|MINimum|MAXimum
Setting the level tone reference threshold. Based on this value, the level scale is zoomed in the range
-15 to +15 dB and the level magnitude is output as a tone (0 dB corresponds to approx. 400 Hz)
Parameters:
<numeric_value> level tone reference threshold in dBuV
UP|DOWN increase|decrease of level tone reference threshold by the value
set in the OUTPut:TONE:THReshold:
STEP[:INCRement] command.
MINimum|MAXimum setting the lowest/highest level tone reference threshold
*RST state:
15 dBuV
Example:
OUTPut:TONE:THReshold 35 dBuV