9
DOC0211
eRev. 1.08/15/2023
User Guide
CTechFlowVPX® System
The power supply must be properly certified for the location and installed according to the CEC and NEC wiring methods
accepted to the Local Authority Having Jurisdiction (LAHJ).
Installation Instructions (in C1D2 Hazardous Locations)
The connectors used shall be capable of being connected to one of the wiring methods as permitted by the NEC for the US or CE
Code, Part I, for Canada for the hazardous location in accordance with any applicable restrictions including issues such as the
routing, support, and maximum length.
Note: See the following United States and Canadian Code references for additional details on some permitted wiring methods for
Class I locations:
a) Sections 501.105(B)(6) or 501.140 of the NEC or Rule J18-160 of the CE Code, Part I for Extra - hard usage cord;
b) Section 501.105(B) of the NEC for Power-limited tray cable (Types PLTC or PLTC-ER), Instrumentation tray cable (Types ITC
or ITC-ER) and Tray cable (Types TC or TC-ER)) ; and
c) Rule J18-152 of the CE Code, Part I for Tray cable (Type TC), Armoured cable with overall nonmetallic jacket (TECK90),
Control and Instrumentation cable (Type ACIC) and Non-armoured control and instrumentation cable (Type CIC).
In the United States, cable assemblies and the associated separate plugs and sockets referenced in 8.8.1, shall be in accordance with
UL 2238 or UL 2237. In Canada, external connectors shall be in accordance with CSA C22.2 No. 159. Alternatively, cable assemblies
and the associated separate plugs and sockets may be used that comply with other relevant standards provided these other
standards include requirements that address the following:
a) The voltage and current rating of the cable assemblies and the associated separate plugs and sockets.
b) For field wiring applications, the suitability of the cable assemblies and the associated separate plugs and sockets for field
installation.
c) Use of the cable assemblies and the associated separate plugs and sockets with one or more of the permitted wiring
methods.
Note: Connectors evaluated to UL 1977 are not suitable for field wiring applications.
IP65 Certification
The FlowVPX instrument has been tested and certified for IP65 compliance according to IEC 60659. Testing was conducted with the
FlowVPX System fully assembled with Detector and Flow Cell with fresh seals installed. C Technologies document CTech FlowVPX
IP65 Declaration of Conformity CTIDC-0001 is included in shipment.
The Beams System has been tested and certified for IP65 compliance according to IEC 60659. Testing was conducted with the Beams
System in any configuration, fully assembled or on its own. The Declaration of Conformity, CTech Beams System CTIDC-0005 is
included in shipment.
WEEE Compliance
Repligen Corporation has met its obligations to the EU WEEE and Battery Directive by registering in those countries to which
Repligen Corporation is an importer. Repligen Corporation has also elected to join WEEE and Battery Compliance Schemes in some
countries to help manage customer returns at end-of-life.
CAUTION: Testing and certification does not include the FlowVPX power supply, which must be installed outside of the
hazardous location, or otherwise isolated from the hazardous environment.
WARNING: Explosion hazard for hazardous locations. Do not connect or disconnect any cabling while energized.