Page #105 background imageLoading...
Page #105 background image
RIGOL
Users Guide for DM3058
3-11
Example 7: Reading Hold
Reading Hold can help user to obtain a stable reading and hold it on the display of
Front Panel. The reading would be hold all the same although testing pen had been
withdrew.
Operating Steps:
1. Connecting the red lead (Test leads) to Terminal Input-HI (tested circuit) and
black lead to Terminal Input-LO.
2. Press to start DC Voltage Measurement and choose an appropriate
range.
3. Set Parameters of Holding Trigger.
(1) Press Auto Hold On/Off 0.1% Holding range:
0.1%
Figure 3-27 Reading holding interface
(2) Press to save all changes.
(3) Repress to finish setting and close menu.
4. Press two times to start Hold function.
5. Connecting Test leads into circuit and start measuring.

Table of Contents

Other manuals for Rigol DM3058

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DM3058 and is the answer not in the manual?

Rigol DM3058 Specifications

General IconGeneral
Diode testYes
Product typeDigital multimeter
Frequency range0.02 - 1000 kHz
AC current range0.02 - 10 A
AC voltage range0.2 - 750 V
DC current range0.0002 - 10 A
DC voltage range0.2 - 1000 V
Resistance range200 - 100000000 Ω
Capacitance range2 - 10000000 nF
Measuring categoryCAT III 300V
Basic accuracy (frequency)0.01 + 0.003
Basic accuracy (DC current)0.055 + 0.005
Basic accuracy (DC voltage)0.015 + 0.003
Basic accuracy (resistance)0.020 + 0.003
Basic accuracy (capacitance)1 + 0.5
Display digits5.5 digits
Display size (HxV)256 x 64 mm
Storage temperature (T-T)-20 - 70 °C
Operating temperature (T-T)0 - 50 °C
Weight and Dimensions IconWeight and Dimensions
Depth107 mm
Width231.6 mm
Height290.5 mm
Weight2500 g