Users Guide for DM3058
5-3
Electric Technique Characteristics
DC Characteristics
Accuracy Specifications (% of reading + % of range)
[1]
Function
Range
[2]
Test Current
or Burden
Voltage
1 Year
23±5
Temperature
Coefficient
0 to 18
28 to 55
DC
Voltage
200.000mV
0.015 + 0.004
0.0015+0.0005
2.00000V
0.015 + 0.003
0.0010+0.0005
20.0000V
0.015 + 0.004
0.0020+0.0005
200.000V
0.015 + 0.003
0.0015+0.0005
1000.00V
[4]
0.015 + 0.003
0.0015+0.0005
DC
Current
[8]
200.000uA
<8mV
0.055 + 0.005
0.003+0.0010
2.00000mA
<80mV
0.055 + 0.005
0.002+0.001
20.0000mA
<0.05V
0.095 + 0.020
0.008+0.001
200.000mA
<0.5V
0.070 + 0.008
0.005+0.001
2.00000A
<0.1V
0.170 + 0.020
0.013+0.001
10.0000A
[5]
<0.3V
0.250 + 0.010
0.008+0.001
Resistance
[3]
200.000Ω
1mA
0.030 + 0.005
0.0030+0.0006
2.00000kΩ
1mA
0.020 + 0.003
0.0030+0.0005
20.0000kΩ
100uA
0.020 + 0.003
0.0030+0.0005
200.000kΩ
10uA
0.020 + 0.003
0.0030+0.0005
2.00000MΩ
1uA
0.040 + 0.004
0.0040+0.0005
10.0000MΩ
200nA
0.250 + 0.003
0.0100+0.0005
100.000MΩ
200nA ||
10MΩ
1.75 + 0.004
0.2000+0.0005
Diode Test
2.0000V
[6]
1mA
0.05 + 0.01
0.0050+0.0005
Continuity
2000Ω
1mA
0.05 + 0.01
0.0050+0.0005
NOTE:
[1] Specifications are for 30 minute warmup and select resolving index to 5 3/4 and calibration
temperature 18 ~ 28 .
[2] 20% over range on all ranges, except DCV 1000V, ACV 750V, DCI and ACI 10A range.
[3] Specifications are for 4wire resistance function, or 2wire resistance using P/F operation.
Without P/F operation, add 0.2 Ω additional errors in 2wire resistance function.
[4] For each additional volt over ± 500 VDC add 0.02 mV of error.
[5] For current terminal, > 7A DC or ACRMS for 30 seconds ON and 30 seconds OFF.
[6] Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test
current is typical. Variation in the current source will create some variation in the voltage
drop across a diode junction.

Table of Contents