User’s Guide for DM3058
Auto Triggering
Auto triggering takes continuous measurement at the highest sample rate for the
specified measurement configuration (function, range, resolution, and so forth).
Auto trigger is the default trigger mode at system power up.
Press Auto, the display shows:
Figure 2-67 Setting interface of Auto Trigger
Table 2-22 Auto Trigger Function Menu Description
1. Interval
The interval means the waiting time before sampling since trigger signal has
been sent. The interval is different under different measure speed.
Fast: the interval default is 8ms, setting range is 8ms~2000ms.
Middle: the interval default is 50ms, setting range is 50ms~2000ms.
Slow: the interval default is 400ms, setting range is 400ms~2000ms.
The default measure speed is slow, so the default interval is 400ms.
The interval is stored in volatile storage.
2. Reading Hold
The reading hold mode captures and holds a reading on the front panel display
when a stable reading is detected. The system Hold ranges are 0.01%, 0.1%,
1%, and 10%.
Set interval time in 400~2000ms.
Set turn the Reading Hold function ON or OFF.
Save all changes, back to a higher level menu.