RIGOL
User’s Guide for DM3058
Secondly Function
is used to open Double Display function and enter into save mode of
configuration or parts of shortcut keys. After pressing , this button will shine
and the instrument is going to wait for the secondly operation.
Double Display can show two properties for one signal. Stat. function can be used
in Main Display.
Auto Range function is adopted by Vice Display. If the same range is used by both
Main and Vice Display, Stat. function will be unavailable for Vice Display, meanwhile
the relative measurement, db, dBm and data cannot be saved into “History”.
Press and Measure one by one.
Table 2-13 Available Main/Vice Function Combination
Explanations:
If the same measurement function is used by both Main and Vice Display, then
it will measure only one time and update the display.
If Math function is used in Main Display, the result will be shown in Main
Display and the measurement value before math is shown in Vice Display.
If Sensor is used in Main Display, and both of Display has same type, then the
result will be shown in Main Display and the measurement value is shown in
Vice Display.
If different measurement function is used, the results will be shown in
corresponding display.