RIGOL
User’s Guide for DM3058
Table 2-5 Measurement Characteristics of Resistance
200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ, 10MΩ, 100MΩ
1000V on every range (HI Terminal)
NOTE*: All the ranges enable to obtain 20% ranges higher than original; besides
Both Manual and Auto are available for setting every range.
4. Set Relative Value ( Optional)
Press REL to open or close Relative math function. When it is open, “REL” is
shown over the screen and the reading displayed is a relative value which
comes from the result of Actual measurement vale reduces selected value.
(For the details about how to set, please refer to” Math” in Chapterer2.)
5. Select required measurement ratio by using Left or Right direction keys and
read the measurement result.
6. View measured history data
Press History to enter into the following interface to check the data getting
from this measurement and save.
Figure 2-19
View interface of history information
There are three types of way for viewing historical date: “Info”,” List” and “HistoG”.
Press Save after checking, by pressing Update, the History information will be
refreshed to be current newest information.
NOTE
You are suggested to make use of Relative function when measuring
small resistance to reduce or escape impedance error from Test leads.