Users Guide for DM3058
2-17
Table 2-6 Measurement Characteristics of 4-Wire Resistance
Ranges*
200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ, 10MΩ, 100MΩ
Open-circuit
Voltage
<8V
Import Protection
(1). 1000V on each range (HI Terminal)
(2). 200V on each range (HI Sense, LO Sense)
Configurable
Parameters
Range, Null value
NOTE*: All the ranges enable to obtain 20% ranges higher than original; besides,
Both Manual and Auto are available for setting every range.
4. Set Relative Value ( Optional)
Press REL to open or close Relative math function. When it is open, “RELis
shown over the screen and the reading displayed is a relative value which
comes from the result of Actual measurement vale reduces selected value.
(For the details about how to set, please refer to” Math” in Chapterer2.)
5. Select required measurement ratio by using Left or Right direction keys and
read the measurement result.
6. View History data
7. Use history function.
Press History to enter into the following interface and check the data getting
from this measurement and save.
Figure 2-22 History Information
There are three types of way for viewing historical date: “Info”, List” and HistoG.
Press Save after checking. By pressing Update, the History information will be
refreshed to be current newest information.