RIGOL  Chapter 5 To Trigger the Oscilloscope 
5-48    DS2000E User’s Guide 
Trigger Output Connector 
 
The trigger output connector ([Trigger Out]) on the rear panel of the oscilloscope 
can output trigger signals determined by the current setting. 
 
 
 
The oscilloscope can output a signal that reflects the current capture rate of the 
oscilloscope from the [Trigger Out] connector each time a trigger is generated. 
Connect the signal to a waveform display device and measure the signal frequency; 
the measurement result is the same with the current capture rate. 
 
Note: If Utility  AuxOutput is set to "PassFail" or Utility  Pass/Fail  
AuxOutput is set to "ON", during the pass/fail test, the oscilloscope will output a 
negative pulse through the [Trigger Out] connector each time a failure is detected; 
the oscilloscope will output low level continuously through the [Trigger Out] 
connector when no failure is detected.  
 
 
Trigger Output Connector