Chapter 1 Quick Start RIGOL
DS2000E Users Guide 1-7
Probe Compensation
When the probe is used for the first time or when the probe compensation signal
does not match that shown in Figure 1-7, you should compensate the probe to match
the input channel of the oscilloscope. The non-compensated or poorly compensated
probe may cause measurement errors. The probe compensation procedures are as
follows:
1. Perform Step 1, 2, 3, and 4 in "Function Inspection".
2. Check the displayed waveforms and compare them with the following figures.
Figure 1-8 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the waveform is displayed as "Perfectly compensated" in
the figure above.
Over compensated Perfectly compensated Under compensated