Chapter 2 To Set the Vertical System    RIGOL 
DS2000E User’s Guide    2-5 
Input Impedance 
 
This oscilloscope provides two input impedance modes (1 MΩ (default) and 50 Ω) to 
reduce the circuit load caused by the interaction of the oscilloscope and the circuit to 
be tested.  
  1MΩ: the input impedance of the oscilloscope is rather high and the current 
flowing into the oscilloscope from the circuit under test can be ignored.  
  50Ω: match the oscilloscope with a device with 50 Ω output impedance.   
 
Press CH1  Input to set the input impedance of the oscilloscope. When "50Ω" is 
selected, "Ω" is displayed in the channel status label at the bottom of the screen. 
 
 
Note: The setting of the input impedance affects the settable range of the vertical 
scale and vertical offset of the corresponding channel. For the details, please refer to 
"Specifications". 
 
 
Waveform Invert 
 
Press CH1  Invert to turn waveform invert on or off. When waveform invert is 
turned off, the waveform display is normal; when waveform invert is turned on, the 
waveform voltage values are inverted (as shown in the figures below).  
(a) "Invert" is Off 
(b) "Invert" is On  
Figure 2-1 Waveform Invert