Chapter 6 MATH and Measurements    RIGOL 
DS2000E User’s Guide    6-23 
Measurement Setting 
 
1.  Source Selection 
Press Measure  Source to select the source channel (CH1, CH2, or MATH) of 
the time, voltage or area parameter. The color of the parameter icons under 
MENU at the left of the screen will change with the source selected.
 
 
2.  Measurement Range 
Press Measure  Range  Range to select "Screen" or "Cursor" for 
measurement.  
 
When "Cursor" is selected, two cursor lines appear on the screen. At this point, 
press CursorA and CursorB and use   to adjust the position of the two 
cursor lines respectively so as to determine the measurement range. Or, press 
CursorAB and use   to adjust the positions of cursor A and cursor B at the 
same time.  
Note: You can press   continuously to switch among CursorA, CursorB and 
CursorAB. 
 
Figure 6-6 Cursor Region 
 
3.  Delayed Measurement Setting 
Specify source A and source B in the measurement items "Dly A
→B ", "Dly 
A
→B ", "Dly A →B " and "Dly A →B ". 
Press Measure  Setting  Type  "Delay" and then press SourceA and 
SourceB to set the two channel sources (CH1 or CH2) of the delayed 
measurement respectively.