e.
After Measuring, the result is displayed
as showed in the right graph. ※2)
A:
To determine the backsight point only by
F1,press [ENT] to end measuring.
B1:
To determine the backsight point by
F2,press [F2] softkey. As showed in
B1.
B2:
To go directly to the Face-2
measurement after taking a distance
measurement to the BS on Face-1, flip
the telescope. ※3)
Sight the backsight point, press [MSR1]
or [MSR2] to start F2 measurement,
press [ENT] after measuring. If no need
to measure, just press [ENT]. ※2)
f.
Press [DSP] to display a QA screen.
(Quality Assessment)
To record a CP record which stores the
averaged HA, VA, and SD from the
F1/F2 data, press the
[CP] softkey. To record only the ST and
F1/F2 records, without a CP record,
press the [OK] softkey. Press [Abrt] to
return to procedure
g.
Procedure records the station and raw
data to current job and finish setting up
station. Screen returns to BMS.
※1) AZ:Azimuth calculated by coordinates.
※2) Press [▼] or [DSP] to switch QA screen (dHD/dVD)。
dHD/dVD: indicates the difference between the measured distance and the distance calculated from the
known coordinates.
※3) The instrument automatically detects F1/F2.