EasyManua.ls Logo

Roche cobas b 221 system

Roche cobas b 221 system
368 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Roche Diagnostics May 2009
Service Manual · Version 9.0 D-107
cobas b 221 system 7 Troubleshooting
Important test routines
Status not ok :
If an error value is displayed on the halogen lamp x-peak, it indicates a grid shift
(photodiode array movement). If necessary replace the polychromator bilirubin.
e
This error pattern corresponds to Flag no. 6032 "Halogen lamp not OK" on page D-83.
Lamp test – Neon lamp: Activate Neon lamp (green switch). When the test is complete, the values and the
status (ok / not ok ) are displayed in the table.
Status not ok :
The light output of the neon lamp is too low.
If an error value is displayed on the neon lamp x-peak, it indicates a grid shift
(photodiode array movement). If necessary replace the polychromator bilirubin.
e
This error pattern corresponds to Flag no. 6031 "Neon lamp not OK" on page D-83.
Optical status: Use this function to check the polychromator.
When the test is complete, the values and the status (ok / not ok ) are
displayed in the table.
Figure D-12 Lamp test (Halogen lamp-green)
Figure D-13 Lamp test (Neon lamp-green)

Table of Contents

Related product manuals