Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
192User Manual 1173.9240.02 ─ 17
Parameters:
<Band> K | A | KA | Q | U | V | E | W | F | D | G | Y | J | USER
Standard waveguide band or user-defined band.
Note: The band formerly referred to as "A" is now named "KA";
the input parameter "A" is still available and refers to the same
band as "KA".
For a definition of the frequency range for the pre-defined bands,
see table 11-2).
*RST: F (90 GHz - 140 GHz)
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:BAND KA
Sets the band to KA (26.5 GHz - 40 GHz).
Manual operation: See "Band" on page 67
[SENSe:]CORRection:CVL:BIAS <BiasSetting>
This command defines the bias setting to be used with the conversion loss table.
Before this command can be performed, the conversion loss table must be selected
(see [SENSe:]CORRection:CVL:SELect on page 195.
This command is only available with option B21 (External Mixer) installed.
Parameters:
<BiasSetting> numeric value
*RST: 0.0 A
Default unit: A
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:BIAS 3A
Manual operation: See "Write to <CVL table name>" on page 64
See "Bias" on page 67
[SENSe:]CORRection:CVL:CATAlog?
This command queries all available conversion loss tables saved in the
C:\r_s\instr\user\cvl\ directory on the instrument.
This command is only available with option B21 (External Mixer) installed.
Usage:
Query only
[SENSe:]CORRection:CVL:CLEAr
This command deletes the selected conversion loss table. Before this command can
be performed, the conversion loss table must be selected (see [SENSe:
]CORRection:CVL:SELect on page 195).