Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
206User Manual 1173.9240.02 ─ 17
Parameters:
<SampleRate> Range: 1 Hz to 10 GHz
*RST: 32 MHz
Example: INP:DIQ:SRAT 200 MHz
Manual operation: See "Input Sample Rate" on page 69
INPut:DIQ:SRATe:AUTO <State>
If enabled, the sample rate of the digital I/Q input signal is set automatically by the con-
nected device.
This command is only available if the optional Digital Baseband Interface (R&S FSW-
B17) is installed.
Parameters:
<State> ON | OFF
*RST: OFF
Manual operation: See "Input Sample Rate" on page 69
11.4.2.5 Setting up Probes
Probes can be connected to the optional BASEBAND INPUT connectors, if the Analog
Baseband interface ( option R&S FSW-B71) is installed.
[SENSe:]PROBe<p>:ID:PARTnumber?............................................................................206
[SENSe:]PROBe<p>:ID:SRNumber?............................................................................... 207
[SENSe:]PROBe<p>:SETup:MODE.................................................................................207
[SENSe:]PROBe<p>:SETup:NAME?............................................................................... 207
[SENSe:]PROBe<p>:SETup:STATe?...............................................................................208
[SENSe:]PROBe<p>:SETup:TYPE?................................................................................ 208
[SENSe:]PROBe<p>:ID:PARTnumber?
Queries the R&S part number of the probe.
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Return values:
<PartNumber> Part number in a string.
Usage: Query only
Configuring the Measurement