EasyManua.ls Logo

Rohde & Schwarz FSW Series - Page 257

Rohde & Schwarz FSW Series
371 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
257User Manual 1173.9240.02 ─ 17
[SENSe:]ADEMod<n>:ZOOM:LENGth............................................................................. 257
[SENSe:]ADEMod<n>:ZOOM:LENGth:MODE...................................................................257
[SENSe:]ADEMod<n>:ZOOM:STARt............................................................................... 257
[SENSe:]ADEMod<n>:ZOOM[:STATe].............................................................................258
[SENSe:]ADEMod<n>:ZOOM:LENGth <Length>
The command allows you to define the length of the time domain zoom area for the
analog-demodulated measurement data in the specified window manually. If the length
is defined manually using this command, the zoom mode is also set to manual.
Parameters:
<Length> *RST: sweep time
Length of the zoom area in seconds.
Example:
ADEM:ZOOM:LENG 2s
Zoom mode is set to manual and the zoom length to 2 seconds.
Manual operation: See "Length" on page 109
[SENSe:]ADEMod<n>:ZOOM:LENGth:MODE <Mode>
The command defines whether the length of the zoom area for the analog-demodula-
ted measurement data is defined automatically or manually in the specified window.
Parameters:
<Mode> AUTO | MAN
AUTO
(Default:) The number of sweep points is used as the zoom
length.
MAN
The zoom length is defined manually using [SENSe:
]ADEMod<n>:ZOOM:LENGth.
*RST: AUTO
Example:
ADEM:ZOOM:LENG:MODE MAN
Zoom function uses the length defined manually.
Manual operation: See "Length" on page 109
[SENSe:]ADEMod<n>:ZOOM:STARt <Time>
The command selects the start time for the zoomed display of analog-demodulated
measurements in the specified window. The maximum possible value depends on the
measurement time, which is set and can be queried with the [SENSe:]ADEMod:
MTIMe command.
If the zoom function is enabled, the defined number of sweep points are displayed from
the start time specified with this command.
Configuring the Measurement

Table of Contents

Other manuals for Rohde & Schwarz FSW Series

Related product manuals