Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
240User Manual 1173.9240.02 ─ 17
Example:
DISP:TRAC:Y:SPAC LIN
Selects linear scaling in %.
Usage: SCPI confirmed
Manual operation: See "Deviation" on page 119
See "Scaling" on page 120
11.4.6 Configuring Data Acquisition
The following remote commands are required to configure which data is to be acquired
and then demodulated in a remote environment.
MSRA/MSRT operating mode
In MSRA/MSRT operating mode, only the MSRA/MSRT Master channel actually cap-
tures data from the input signal. The data acquisition commands for the Analog
Demodulation application in MSRA/MSRT mode define the analysis interval.
For details on the MSRA operating mode see the R&S FSW MSRA User Manual. For
details on the MSRT operating mode see the R&S FSW Realtime Spectrum Applica-
tion and MSRT Operating Mode User Manual.
[SENSe:]ADEMod:MTIMe...............................................................................................240
[SENSe:]ADEMod:RLENgth?.......................................................................................... 240
[SENSe:]ADEMod:SET...................................................................................................241
[SENSe:]ADEMod:SPECtrum:BANDwidth|BWIDth[:RESolution]......................................... 242
[SENSe:]ADEMod:SRATe?.............................................................................................242
[SENSe:]BANDwidth|BWIDth:DEMod...............................................................................242
[SENSe:]BANDwidth|BWIDth:DEMod:TYPE..................................................................... 243
[SENSe:]BANDwidth|BWIDth[:RESolution]....................................................................... 243
[SENSe:]SWEep:COUNt.................................................................................................243
[SENSe:]SWEep:POINts.................................................................................................244
[SENSe:]ADEMod:MTIMe <Time>
This command defines the measurement time for analog demodulation.
Parameters:
<Time> *RST: 62.5us
Example:
ADEM:MTIM 62.5us
Sets the measurement time to 62.5 µs.
Manual operation: See "Measurement Time (AQT)" on page 103
[SENSe:]ADEMod:RLENgth?
This command returns the record length set up for the current analog demodulation
measurement.
Configuring the Measurement