Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
258User Manual 1173.9240.02 ─ 17
Parameters:
<Time> Range: 0 s to (measurement time – zoom length)
*RST: 0 s
Example:
ADEM:ZOOM:STAT ON
Switches on the zoom function
ADEM:ZOOM:STAR 500us
Sets the starting point of the display to 500 µs.
Manual operation: See "Start" on page 109
[SENSe:]ADEMod<n>:ZOOM[:STATe] <State>
The command enables or disables the time domain zoom function for the analog-
demodulated measurement data in the specified window.
If the zoom function is enabled, the defined number of sweep points are displayed from
the start time specified with [SENSe:]ADEMod<n>:ZOOM:STARt on page 257.
If the zoom function is disabled, data reduction is used to adapt the measruement
points to the number of points available on the display.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
ADEM:ZOOM ON
Switches on the zoom function
Manual operation: See "State" on page 109
11.4.8.3 Configuring the Demodulation Spectrum
The demodulation spectrum defines which span of the demodulated data is evaluated.
● AF evaluation........................................................................................................ 258
● RF evaluation........................................................................................................260
AF evaluation
These settings are only available for AF Spectrum evaluations, not in the time domain.
[SENSe:]ADEMod:AF:CENTer........................................................................................ 258
[SENSe:]ADEMod:AF:SPAN........................................................................................... 259
[SENSe:]ADEMod:AF:SPAN:FULL.................................................................................. 259
[SENSe:]ADEMod:AF:STARt.......................................................................................... 259
[SENSe:]ADEMod:AF:STOP........................................................................................... 259
[SENSe:]ADEMod:AF:CENTer <Frequency>
This command sets the center frequency for AF spectrum result display.
Parameters:
<Frequency> *RST: 1.25 MHz
Configuring the Measurement