EasyManua.ls Logo

Rohde & Schwarz FSW Series - Page 259

Rohde & Schwarz FSW Series
371 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
259User Manual 1173.9240.02 ─ 17
Manual operation: See "AF Center" on page 111
[SENSe:]ADEMod:AF:SPAN <Span>
This command sets the span (around the center frequency) for AF spectrum result dis-
play.
The span is limited to DBW/2 (see [SENSe:]BANDwidth|BWIDth:DEMod
on page 242).
Parameters:
<Span> *RST: 9 MHz
Example:
ADEM:AF:SPAN 200 kHz
Sets the AF span to 200 kHz
Manual operation: See "AF Span" on page 111
[SENSe:]ADEMod:AF:SPAN:FULL
This command sets the maximum span for AF spectrum result display.
The maximum span corresponds to DBW/2 (see [SENSe:]BANDwidth|BWIDth:
DEMod on page 242).
Example:
ADEM:BAND 5 MHz
Sets the demodulation bandwidth to 5 MHz
ADEM:AF:SPAN:FULL
Sets the AF span to 2.5 MHz
Manual operation: See "AF Full Span" on page 112
[SENSe:]ADEMod:AF:STARt <Frequency>
This command sets the start frequency for AF spectrum result display.
Parameters:
<Frequency> *RST: 0 MHz
Example:
ADEM:AF:STAR 0 kHz
Sets the AF start frequency to 0 kHz
ADEM:AF:STOP 500 kHz
Sets the AF stop frequency to 500 kHz
Manual operation: See "AF Start" on page 111
[SENSe:]ADEMod:AF:STOP <Frequency>
This command sets the stop frequency for AF spectrum result display.
Parameters:
<Frequency> *RST: 9 MHz