Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
328User Manual 1173.9240.02 ─ 17
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST: ABSolute
Manual operation: See "X-Axis" on page 152
CALCulate:
LIMit<k>:UPPer:OFFSet <Offset>
This command defines an offset for a complete upper limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Parameters:
<Offset> Numeric value.
*RST: 0
Default unit: dB
Manual operation: See "Y-Offset" on page 149
CALCulate:LIMit<k>:UPPer:SHIFt <Distance>
This command moves a complete upper limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Parameters:
<Distance> Defines the distance that the limit line moves.
The unit depends on CALCulate:LIMit<k>:UNIT
on page 327.
Usage: Event
Manual operation: See "Shift y" on page 153
CALCulate:LIMit<k>:UPPer:SPACing <InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Parameters:
<InterpolType> LINear | LOGarithmic
*RST: LIN
Manual operation: See "X-Axis" on page 152
See "Y-Axis" on page 152
Analyzing Results