Analysis
R&S
®
FSW-K7
152User Manual 1173.9240.02 ─ 17
Margin
Defines a margin for the limit line. The default setting is 0 dB (i.e. no margin).
Remote command:
CALCulate:LIMit<k>:LOWer:MARGin on page 325
CALCulate:LIMit<k>:UPPer:MARGin on page 327
X-Axis
Describes the horizontal axis on which the data points of the limit line are defined.
Includes the following settings:
Domain:
"Hz": for frequency domain
"s": for time domain
Scaling mode: absolute or relative (Hz/s/%) values
For relative values, the frequencies are referred to the currently set center fre-
quency. In the zero span mode, the left boundary of the diagram is used as the ref-
erence.
Scaling: linear or logarithmic
Remote command:
CALCulate:LIMit<k>:LOWer:SPACing on page 326
CALCulate:LIMit<k>:UPPer:SPACing on page 328
CALCulate:LIMit<k>:LOWer:MODE on page 325
CALCulate:LIMit<k>:UPPer:MODE on page 327
CALCulate:LIMit<k>:CONTrol:DOMain on page 323
Y-Axis
Describes the vertical axis on which the data points of the limit line are defined.
Includes the following settings:
Level unit
Scaling mode: absolute or relative (dB/%) values
Relative limit values refer to the reference level.
Limit type: upper or lower limit; values must stay above the lower limit and below
the upper limit to pass the limit check
Remote command:
CALCulate:LIMit<k>:UNIT on page 327
CALCulate:LIMit<k>:LOWer:SPACing on page 326
CALCulate:LIMit<k>:UPPer:SPACing on page 328
Data points
Each limit line is defined by a minimum of 2 and a maximum of 200 data points. Each
data point is defined by its position (x-axis) and value (y-value). Data points must be
defined in ascending order. The same position can have two different values.
Remote command:
CALCulate:LIMit<k>:CONTrol[:DATA] on page 323
CALCulate:LIMit<k>:LOWer[:DATA] on page 324
CALCulate:LIMit<k>:UPPer[:DATA] on page 327
Limit Line Settings and Functions