Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
325User Manual 1173.9240.02 ─ 17
CALCulate:LIMit<k>:LOWer:MARGin <Margin>
This command defines an area around a lower limit line where limit check violations
are still tolerated.
Parameters:
<Margin> numeric value
*RST:  0
Default unit: dB
Manual operation: See "Margin" on page 152
CALCulate:
LIMit<k>:LOWer:MODE <Mode>
This command selects the vertical limit line scaling.
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:  ABSolute
Manual operation: See "X-Axis" on page 152
CALCulate:LIMit<k>:LOWer:OFFSet <Offset>
This command defines an offset for a complete lower limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Parameters:
<Offset> Numeric value.
*RST:  0
Default unit: dB
Manual operation: See "Y-Offset" on page 149
CALCulate:LIMit<k>:LOWer:SHIFt <Distance>
This command moves a complete lower limit line vertically.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Parameters:
<Distance> Defines the distance that the limit line moves.
The unit depends on CALCulate:LIMit<k>:UNIT
on page 327.
Analyzing Results