Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
324User Manual 1173.9240.02 ─ 17
Parameters:
<Offset> Numeric value.
The unit depends on the scale of the x-axis.
*RST: 0
Manual operation: See "X-Offset" on page 149
CALCulate:LIMit<k>:CONTrol:SHIFt <Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Parameters:
<Distance> Numeric value.
The unit depends on the scale of the x-axis.
Manual operation: See "Shift x" on page 153
CALCulate:LIMit<k>:CONTrol:SPACing <InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from one horizontal point to the next.
Parameters:
<InterpolMode> LINear | LOGarithmic
*RST: LIN
Example:
CALC:LIM:CONT:SPAC LIN
CALCulate:LIMit<k>:LOWer[:DATA] <LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Parameters:
<LimitLinePoints> Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with CALCulate:
LIMit<k>:CONTrol[:DATA]. If not, the R&S FSW either adds
missing values or ignores surplus values.
The unit depends on CALCulate:LIMit<k>:UNIT
on page 327.
*RST: Limit line state is OFF
Usage: SCPI confirmed
Manual operation: See "Data points" on page 152