Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
323User Manual 1173.9240.02 ─ 17
CALCulate:LIMit<k>:CONTrol[:DATA] <LimitLinePoints>
This command defines the horizontal definition points of a limit line.
Parameters:
<LimitLinePoints> Variable number of x-axis values.
Note that the number of horizontal values has to be the same as
the number of vertical values set with CALCulate:LIMit<k>:
LOWer[:DATA] or CALCulate:LIMit<k>:UPPer[:DATA]. If
not, the R&S FSW either adds missing values or ignores surplus
values.
The unit is Hz or s.
*RST: -
Usage: SCPI confirmed
Manual operation: See "Data points" on page 152
CALCulate:
LIMit<k>:CONTrol:DOMain <SpanSetting>
This command selects the domain of the limit line.
Parameters:
<SpanSetting> FREQuency | TIME
*RST: FREQuency
Manual operation: See "X-Axis" on page 152
CALCulate:LIMit<k>:CONTrol:MODE <Mode>
This command selects the horizontal limit line scaling.
Parameters:
<Mode> ABSolute
Limit line is defined by absolute physical values (Hz or s).
RELative
Limit line is defined by relative values related to the center fre-
quency (frequency domain) or the left diagram border (time
domain).
*RST: ABSolute
CALCulate:LIMit<k>:CONTrol:OFFSet <Offset>
This command defines an offset for a complete limit line.
Compared to shifting the limit line, an offset does not actually change the limit line defi-
nition points.
Analyzing Results