Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
327User Manual 1173.9240.02 ─ 17
Manual operation: See "Name" on page 151
CALCulate:LIMit<k>:UNIT <Unit>
This command defines the unit of a limit line.
Parameters:
<Unit> If you select dB as the limit line unit, the command automatically
turns the limit line into a relative limit line.
*RST: DBM
Manual operation: See "Y-Axis" on page 152
CALCulate:LIMit<k>:UPPer[:DATA] <LimitLinePoints>
This command defines the vertical definition points of an upper limit line.
Parameters:
<LimitLinePoints> Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with CALCulate:
LIMit<k>:CONTrol[:DATA]. If not, the R&S FSW either adds
missing values or ignores surplus values.
The unit depends on CALCulate:LIMit<k>:UNIT
on page 327.
*RST: Limit line state is OFF
Usage: SCPI confirmed
Manual operation: See "Data points" on page 152
CALCulate:LIMit<k>:UPPer:MARGin <Margin>
This command defines an area around an upper limit line where limit check violations
are still tolerated.
Parameters:
<Margin> numeric value
*RST: 0
Default unit: dB
Manual operation: See "Margin" on page 152
CALCulate:LIMit<k>:UPPer:MODE <Mode>
This command selects the vertical limit line scaling.
Analyzing Results