Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
329User Manual 1173.9240.02 ─ 17
CALCulate:LIMit<k>:UPPer:STATe <State>
This command turns an upper limit line on and off.
Before you can use the command, you have to select a limit line with CALCulate:
LIMit<k>:NAME on page 326.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: SCPI confirmed
Manual operation: See "Visibility" on page 149
CALCulate:LIMit<k>:UPPer:THReshold <Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S FSW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Parameters:
<Limit> Numeric value.
The unit depends on CALCulate:LIMit<k>:UNIT
on page 327.
*RST: -200
Default unit: dBm
Manual operation: See "Threshold" on page 151
11.8.2.2 Managing Limit Lines
Useful commands for managing limit lines described in the R&S FSW User Manual:
●
MMEM:SEL[:ITEM]:LIN:ALL
●
MMEM:STOR:TYPE
●
MMEM:LOAD:TYPE
Remote commands exclusive to managing limit lines:
CALCulate:LIMit:ACTive?............................................................................................... 329
CALCulate:LIMit<k>:COPY............................................................................................. 330
CALCulate:LIMit<k>:DELete............................................................................................330
CALCulate:LIMit<k>:STATe............................................................................................ 330
CALCulate:LIMit<k>:TRACe<t>:CHECk........................................................................... 331
CALCulate:LIMit:ACTive?
This command queries the names of all active limit lines.
Analyzing Results