Instrument Functions
NRP2
136User Manual 1173.9140.02 ─ 07
Fig. 4-62: Statistics display mode with two traces (AWGN)
From the gathered samples, four parameters for each trace are indicated in the info
panel:
●
the peak power
●
the average power
●
the measurement result at the position of the power (x) marker
●
the power value related to the result (y) marker
In case of AWGN only the measurement result at the position of the power (x) marker
is indicated.
Only in the case of a non-synchronized aquisition interval ("Cont" mode), the indicated
average power value relates to the complete signal. Otherwise, the average power
within the selected gate is indicated.
4.5.8.1
Controls in statistics display mode
Display
Selects another view, see chapter 4.5.2, "Trace window", on page 114.
Config Grid...
Activates the additional tabular display of values or opens the subdialog to configure
the diagram and to numerically set the marker positions, see chapter 4.5.8.2, "Statis-
tics diagram settings", on page 137.
Pos Scale
Defines if the the cursor keys position or scale the displayed curve(s).
Note: There are restrictions on the use of these controls. The vertical scale ranges
from 0% to 100% maximum for all statistic functions except for PDF. Similar restric-
tions apply to the horizontal axis, depending on the power range of the sensors.
"Pos"
The cursor keys shift the curve(s) in horizontal and vertical direction
"Scale"
The cursor keys expand and compress the curve(s).
(marker)
Selects a marker for positioning by means of the cursor keys.