Remote Control - Commands
NRP2
226User Manual 1173.9140.02 ─ 07
CALCulate<[1]..8>:EXTRemes:RESet
The R&S NRP2 stores the maximum and minimum values for each calculate block.
The current measured value is stored with CALC[1..8]:EXTR:RES
as the new mini-
mum and maximum value. The extreme values can be queried with
CALC[1..8]:MIN:DATA? and CALC[1..8]:MAX:DATA?.
Usage: Event
CALCulate<[1]..8>:FEED<[1]|2> <feed>
The sensors connected with the calculate block via the primary and secondary channel
can yield differently processed data: peak value, average value, peak-to-average value
or even a measurement sequence, as is the case in the Trace mode. The CALC:FEED
determines which of this data is to be processed by the calculate block. Depending on
the measurement mode, the following settings are possible:
Table 6-9: <string> values
Primary
channel
measure-
ment mode
<string> Meaning
ContAv
"POWer:AVERage"
"POWer:PEAK"
"POWer:RANDom"
Average value
Peak value
Randomly selected value from the measurement
interval
Burst
"POWer:AVERage"
"POWer:PEAK"
"POWer:RANDom"
Average value
Peak value
Randomly selected value from the measurement
interval
Timeslot
"POWer:AVERage"
"POWer:PEAK"
"POWer:RANDom"
Average value
Peak value
Randomly selected value from the measurement
interval
Timegate
"POWer:AVERage"
"POWer:PEAK"
"POWer:RANDom"
Average value
Peak value
Randomly selected value from the measurement
interval
Trace
"POWer:TRACe"
"POWer:PEAK:TRACe"
"POWer:RANDom:
TRACe"
"POWer:AVERage ON SWEep[1..4]"
"POWer:PTAVerage ON
SWEep[1..4]"
"POWer:PEAK ON SWEep[1..4]"
Measurement sequence
Peak value of the samples per test point
Randomly selected value
Average value in time gate 1, 2, 3 or 4
Peak-to-average value in time gate 1, 2, 3 or 4
Peak value in time gate 1, 2, 3 or 4
Statistic
"CCDF:TRACe"
"PDF:TRACe"
CCDF
PDF